This post is a continuation of my previous post on Permanent magnet moving coil instruments . This post is written since a is request made by the face book user Shivany Sweet
Torque Equation for PMMC
The equation for the delevoped torque of the PMMC can be obtained from the basic law of electromagnetic torque. The deflecting torque is given by,
Td = NBAI
Where,
Td = deflecting torque in N-m
B = flux density in air gap, Wb/m2
N = Number of turns of the coils
A = effective area of coil m2
I = current in the moving coil, amperes
Therefore, Td = GI
Where, G = NBA = constant
The controlling torque is provided by the springs and is proportional to the angular deflection of the pointer.
Tc = KØ
Where,
Tc = Controlling Torque
K = Spring Constant Nm/rad or Nm/deg
Ø = angular deflection
For the final steady state position,
Td = Tc
Therefore GI = KØ
So, Ø = (G/K)I or I = (K/G) Ø
Thus the deflection is directly proportional to the current passing through the coil. The pointer deflection can therefore be used to measure current.
Showing posts with label PMMC. Show all posts
Showing posts with label PMMC. Show all posts
Thursday, September 15, 2011
Thursday, March 24, 2011
Temperature compensation and Error in PMMC:
The basic pmmc instrument is sensitive to the temperature. The magnetic field strength and spring tension decrease with increase in temperature. The coil resistance increases with increase in the temperature. Thus pointer reads low for a given current. The meter tends to read low by approximately 0.2% per Celsius rise in the temperature. Hence the temperature compensation is provided by appropriate use of series and shunt resistance of copper and manganin.
The simple temperature compensation circuit for PMMC uses a resistance in series with a movable coil, as shown in the figure. The resistor is called swamping resistor. It is made up of manganin having practically zero temperature coefficients, combined with copper in the ratio of 20/1 or 30/1.
The resultant resistance of the coil and the swamping resistor increases slightly as temperature increases, just enough to compensate the change in springs and magnet due to temperature. Thus the effect of temperature is compensated.
More complicated but complete cancellation of temperature effects can be obtained by using the swamping resistors in series and parallel combination as shown in figure.
In this circuit, by correct proportioning of copper and manganin parts, complete cancellation of the temperature effects can be achieved.
The basic sources of error in PMMC instruments are friction, temperature and aging of various parts. To reduce the frictional errors ratio of torque to weight is made very high.
The most serious errors are produced by the heat generated or by changes in the temperature. This changes the resistance of the working coil, causing large errors. In case of voltmeters, a large series resistance of very low temperature coefficient is used. This reduces the temperature errors.
The aging of permanent magnet and control springs also cause errors. Opposite errors in PMMC is caused by weakening of magnet and spring cause. The weakening of magnet causes less deflection while weakening of control springs cause large deflection, for a particular value of current. The proper use of material and pre-ageing during manufacturing can reduce the errors due to weakening of control springs.
Temperature compensation
The simple temperature compensation circuit for PMMC uses a resistance in series with a movable coil, as shown in the figure. The resistor is called swamping resistor. It is made up of manganin having practically zero temperature coefficients, combined with copper in the ratio of 20/1 or 30/1.
The resultant resistance of the coil and the swamping resistor increases slightly as temperature increases, just enough to compensate the change in springs and magnet due to temperature. Thus the effect of temperature is compensated.
More complicated but complete cancellation of temperature effects can be obtained by using the swamping resistors in series and parallel combination as shown in figure.
In this circuit, by correct proportioning of copper and manganin parts, complete cancellation of the temperature effects can be achieved.
Errors in PMMC Instrument
The basic sources of error in PMMC instruments are friction, temperature and aging of various parts. To reduce the frictional errors ratio of torque to weight is made very high.
The most serious errors are produced by the heat generated or by changes in the temperature. This changes the resistance of the working coil, causing large errors. In case of voltmeters, a large series resistance of very low temperature coefficient is used. This reduces the temperature errors.
The aging of permanent magnet and control springs also cause errors. Opposite errors in PMMC is caused by weakening of magnet and spring cause. The weakening of magnet causes less deflection while weakening of control springs cause large deflection, for a particular value of current. The proper use of material and pre-ageing during manufacturing can reduce the errors due to weakening of control springs.
Wednesday, March 23, 2011
Permanent magnet moving coil instruments (PMMC)
The permanent magnet moving coil instruments are most accurate type for direct current measurements. The action of these instruments is based on the motoring principle. When a current carrying coil is placed in the magnetic field produced by permanent magnet, the coil experiences a force and moves. As the coil is moving and the magnet is permanent, the instrument is called permanent magnet moving coil instrument. This basic principle is called D’Arsonval principle. The amount of force experienced by the coil is proportional to the current passing through the coil.
The PMMC instrument is shown in the below images.
The moving coil is either rectangular or circular in shape. It has number of turns of fine wire. The coil is suspended so that it is free to turn about its vertical axis. The coil is placed in uniform, horizontal and radial magnetic field of a permanent magnet in the shape of a horse-shoe. The iron core is spherical if coil is circular and is cylindrical if the coil is rectangular. Due to iron core, the deflecting torque increase, increasing the sensitivity of the instrument.
The controlling torque is provided by two phosphor bronze hair springs.
The damping torque is provided by eddy current damping. It is obtained by movement of aluminum former, moving in the magnetic field of the permanent magnet.
The pointer is carried by the spindle and it moves over a graduated scale. The pointer has light weight so that it deflects rapidly. The mirror is placed below the pointer to get the accurate reading by removing the parallax. The weight of the instrument is normally counter balanced by the weights situated diametrically opposite and rapidly connected to it. The scale markings of the basic d.c PMMC instruments are usually linearly spaced as the deflecting torque and hence the pointer deflections are directly proportional to the current passing through the coil.
The top view of PMMC instrument is shown in the below image.
Advantages of PMMC
The various advantages of PMMC instruments are,
It has uniform scale.
With a powerful magnet, its torque to weight ratio is very high. So operating current of PMMC is small.
The sensitivity is high.
The eddy currents induced in the metallic former over which coil is wound, provide effective damping.
It consumes low power, of the order of 25 W to 200 mW.
It has high accuracy.
Instrument is free from hysteresis error.
Extension of instrument range is possible.
Not affected by external magnetic fields called stray magnetic fields.
Disadvantages of PMMC
The various disadvantages of PMMC instruments are ,
PMMC is Suitable for direct current measurement only.
Ageing of permanent magnet and the control springs introduces the errors.
The cost is high due to delicate construction and accurate machining.
The friction is due to jewel-pivot suspension.