Saturday, September 29, 2012

SAYBOLT VISCOMETER (EFFLUX CUP VISCOMETERS)

Definition of SAYBOLT VISCOMETER
A device used to measure the viscosity of a fluid. The saybolt viscometer controls the heat of the fluid and the viscosity is the time is takes the fluid to fill a 60cc container.

Efflux cup viscometers are most commonly used for fieldwork to measure the viscosity of oils, syrups, varnish, paints and Bitumen emulsions. The testing procedure is quite similar to the capillary-tube viscometers where efflux time of a specified volume of fluid is measured through fixed orifice at the bottom of a cup to represent the viscosity of the fluid. Since the viscosity of Newtonian liquid are independent of dimensions of viscometer used, it is possible to convert the efflux times to kinematic viscosities by conversion charts or by formulas suggested by the equipment manufacturers.

To obtain high accuracy the liquid holding vessel and orifice are temperature controlled by immersing them in a thermostatically controlled bath. The saybolt viscometer, one of the efflux cup viscometers is the standard instrument for testing petroleum products. There are three types of orifices available-Universal, Furol, Asphalt.

The furol and asphalt orifices, respectively, have an efflux time of approximately, one-tenth and one-hundredth that of the universal orifice. The cup orifice combination should be selected to provide an effllux time within the range of 20 to 100 seconds. Of these types, the universal orifice(saybolt universal viscometer) is most commonly used and its efflux time is designated as saybolt universal seconds(SUS).The universal viscometer measures the time required for 60 cc of sample fluid to flow out through an orifice having dimensions of 0.176 cm in diameter and 1.225 cm in length. Saybolt universal seconds (t) can be converted to kinematic viscosity (v) by the following equations:

When t < 100 secs, v = 0.226t - 195/t Centistokes
When t >100 secs, v = 0.220t - 135/t Centistokes

The viscosity determinations should be conducted in a room free from drafts and rapid changes in temperature the highest degree of accuracy.

Tuesday, August 7, 2012

Electrical Torsion Meter (Torque Measurement using slotted Discs)

Basic principle

Due to the applied torque, there is a relative displacement between the two slotted discs. Due to this relative displacement of the slotted discs, a phase shift exists between the pulse generated by the transducers. When these pulses are connected to an electronic unit, it will show a time lapse between the two pulses. This time lapse between the two pulses is proportional to the twist of the shaft and the torque of the shaft.

Description of Electrical Torsion Meter


  1. The main parts of an electrical torsion meter are as follows:
  2. A shaft connected between a driving engine and a driven load.
  3. Two slotted discs attached on either side of the shaft.
  4. Transducer (magnetic or photo electric ) to count pulses from the slotted disc.


Operation of Electrical Torsion Meter


  1. The teeth produce voltage pulses in the transducers.
  2. When torque is not applied on the shaft, the teeth of the bth the discs perfectly align with each other and hence he voltage pulses produced in the transducers will have zero time difference.
  3. But when torque is applied on the shaft, there is a relative displacement of the slotted discs due to twist experienced by the shaft and hence the teeth of both the discs will not align with each other and hence the voltage pulses produced in the transducer will have a time difference (that is , time lapse).
  4. This time lapse between the pulses of the two discs is proportional to the twist of the shaft and hence the torque of the shaft.
  5. A measure of this time lapse becames of torque when calibrated.


Application of Electrical Torsion Meter


  1. Used to measure torque on rotation shafts.


Advantages of Electrical Torsion Meter


  1. There are no signal leakage problems.
  2. There is no noise creation.



Sunday, July 1, 2012

Seismic – Displacement Sensing Accelerometer

Basic Principle

When a spring – mass – damper system is subjected to acceleration, the mass is displaced, and this displacement of the mass is proportional to the acceleration. Hence a measure of displacement of the mass becomes a measure of acceleration.

Description

The main parts of a seismic accelerometer are as follows:
  1. A seismic mass is suspended from the housing of the accelerometer through a spring.
  2. A damper is connected between the seismic mass and the housing of the accelerometer.
  3. The seismic mass is connected to an electric displacement transducer.

Note: there are two types of seismic – displacement sensing accelerometers namely.
  1. Linear seisimic accelerometer.
  2. Rotational Seismic Accelerometer.
Both of the above mentioned seismic accelerometers work on the same principle and are shown in diagram.

Operation

  1. The accelerometer is fitted on to the structure whose acceleration is to be measured.
  2. Due the acceleration, the seismic mass experience a displacement and this displacement of the mass is proportional to the acceleration.
  3. As the mass is connected to an electric displacement transducer, the output of the transducer depends on the extent – to which the mass is displaced.
  4. Hence the output of the transducer is calibrated to give a direct indication of the acceleration characteristics of the structure.

Tuesday, June 26, 2012

What is a Magnetic Tape Recorders

Before explaining about magnetic tape recorders, I will tell you what a recorder is and what the uses of the recorder are?


A recorder is used to produce a permanent record of the signal that is measured.

A record is used to analyse how one variable varies with respect to another and how the signal saries with time.

The objective of a recording system is to record and preserve information pertaining to measurement at a particular time and also to get an idea of the performance of the unit and to provide the results of the steps taken by the operator.

The basic components of a general recorder are an operating mechanism to position the pen or writer on the paper and a paper mechanism for paper movement and a printing mechanism.

Okay, now you know what is a recorder, why it is used and where it is used. Now I will explain about magnetic tape recorder.

A magnetic tape recorder is used to record data which can be retrieved and reproduced in electrical form again. This recorder can record signals of high frequency.

Description of Magnetic Tape Recorders:


The magnetic tape is made of a thin sheet of tough plastic material; one side of it is coated with a magnetic material (iron oxide). The plastic base is usually polyvinyl chloride (PVC) or polyethylene terephthalate. Recording head, reproducing head and tape transport mechanism are also present.
magnetic tape recorders

Operation of Magnetic Tape Recorders:

  1. The recording head consists of core, coil and a fine air gap of about 10 micrometer. The coil current creates a flux, which passes through the air gap to the magnetic tape and magnetizes the iron oxide particles as they pass the air gap. So the actual recording takes place at the trailing edge of the gap.
  2. The reproducing head is similar to that of a recording head in appearance. The magnetic tape is passes over a reproducing head, thereby resulting in an output voltage proportional to the magnetic flux in the tape, across the coil of the reproducing head. Thus the magnetic pattern in the tape is detected and converted back into original electrical signal.
  3. The tape transport mechanism moves the tape below the head at constant speed without any strain, distrortion or wear. The mechanism much be such as to guide the tape passed by the magnetic heads with great precision, maintain propoer tension and have sufficient tape to magnetic head contact.

Advantages of Magnetic Tape Recorders:

  1. Wide frequency range.
  2. Low distortion.
  3. Immediate availability of the signal in its initial electrical form as no time is lost in processing.
  4. The possibility of erase and reuse of the tape.
  5. Possibility of playing back or reproducing of the recorded signal as many times as required without loss if signal.

Applications of Magnetic Tape Recorders:

  1. Data recording and analysis on missiles, aircraft and satelites.
  2. Communications and spying.
  3. Recording of stress, vibration and analysis of noise.

Monday, June 11, 2012

D’Arsonval Movement (Electrical Analog Meter)

The main aim of a measuring system is to presnt the data in an understandable manner to the user / observer. The data presentation device is called as an output device or terminating device.

The D’Arsonval movement is a current sensing mechanism which is used in DC Ammeter, ohm meter and Voltmeter.

Priniciple of D’Arsonval Movement

When an electric current is passed through a coil placed in a magnetic field, it experiences a force. This force causes a torque in the coil that is fixed to a spindle. The spindle can rotate in fixed bearings.
The rotation of the spindle is proportional to the electric current passed through the coil. This torque that is produced is balanced after a movement against the restoring torques of springs. The torque that is produced that tends to rotate the spindle is termed as D’Arsonval Movement.

Description of D’Arsonval Movement


  1. The arrangement consists of a coil which is wound over an iron core (spindle).
  2. This spindle is place between the two poles of a horse shoe magnet.
  3. The spindle is attached at its end to bearings. Sprial and torsional springs are provided for restrotation of the system when the extraction is removed.
  4. A pointer is attached to the spindle that can sweep over the calibrated scale.


Operation of D’Arsonval Movement


  1. When a current is passed through the coil, it produces a force. Due to this force, a torque is produced in the spindle which rotates it.
  2. When the spindle rotates, it moves a pointer making it sweep over the calibrated scale.
  3. The spring produce a restoring torque. When this restoring torque becomes equal to the excitation torque, the pointer comes to rest.
  4. The rotational movement of the spindle is proportional to the supply (D.C) Current.


Friday, April 13, 2012

Instruments used to measure angular velocity

Angular velocity is measured using devices called tachometers.Tachometers are instruments that measure angular speed by registering anyone of the following

Number of rotations during the contact period (or)

  • Number of rotations per minute (rpm)
  • Tachometers are classified as follows:


Mechanical Tachometers

  1. Revolution counter and timer
  2. Tachoscope
  3. Hand speed indicator
  4. Slipping clutch tachometer
  5. Centrifugal force tachometer
  6. Vibration reed tachometer


Electrical tachometers

  • Drag cup tachometer
  • Commutated capacitor tachometer
  • Tachogenerators
  • DC – Tachogeneraor
  • AC – Tachogeneraor


Contact Less Electrical Tachometers

  1. Inductive pick up tachometer
  2. Stroboscope tachometer
  3. Photoelectric tachometer
  4. Capacitor tachometer



Saturday, April 7, 2012

Devices used to measure dimensions

This post contains a small list of devices used in measuring dimension. They are ,

  1. Steel rules
  2. Vernier calipers
  3. Micrometers
  4. Dial gauge indicator
  5. Gauge blocks
  6. Optical methods (Microscopes, telescopes and optical flats with a monochromatic source of light)
  7. Pneumatic comparators.

Wednesday, January 11, 2012

List of Instruments and Methods used in Level Measurements:

In this post I will list out all the Instruments and Methods used in Level Measurements.


  1. Level measurement using gauge glass technique.
  2. Float type level indication
  3. Float level switch
  4. Float / rope method
  5. Float operated spring loaded level switch
  6. Magnetic float device
  7. Level measurement using displacer and torque tube
  8. Air purge system/bubbler system
  9. Thermal level sensors
  10. Low level conductivity switch
  11. Thermal level sensor with heater and probes
  12. Capsule type thermal level sensor
  13. Expansion type thermal level sensor
  14. Hydrostatic pressure method
  15. Differential pressure method
  16. Level measurement using resistance tapes
  17. Capacitance level detection
  18. Radiometric level detection
  19. Ultrasonic sensors
  20. Microwave time of flight method
  21. Non-conducting optical level sensor
  22. Optical sludge level detector
  23. Light refraction type level switch
  24. Level detection using fiber optics
  25. Radar level transmitters
  26. Laser level devices
  27. Level switches

Thursday, September 15, 2011

Torque Equation for PMMC

This post is a continuation of my previous post on Permanent magnet moving coil instruments . This post is written since a is request made by the face book user Shivany Sweet


Torque Equation for PMMC

The equation for the delevoped torque of the PMMC can be obtained from the basic law of electromagnetic torque. The deflecting torque is given by,

                                                   Td = NBAI

Where,
Td = deflecting torque in N-m
B = flux density in air gap, Wb/m2
N = Number of turns of the coils
A = effective area of coil m2
I = current in the moving coil, amperes

Therefore, Td = GI
Where, G = NBA = constant

The controlling torque is provided by the springs and is proportional to the angular deflection of the pointer.
                                               Tc = KØ

Where,
Tc = Controlling Torque
K = Spring Constant Nm/rad or Nm/deg
Ø = angular deflection
For the final steady state position,
                        Td = Tc
Therefore GI = KØ
So,                      Ø = (G/K)I          or            I = (K/G) Ø

Thus the deflection is directly proportional to the current passing through the coil. The pointer deflection can therefore be used to measure current.

Wednesday, July 27, 2011

How to select a Transducer ?

In a measurement system the transducer is the input element with critical function of transforming some physical quantity to a proportional electrical signal. Selection of the appropriate transducer is therefore the first and perhaps most important step in obtaining accurate results in every instrumentation and control system. A number of elementary questions should be asked before a transducer can be selected, for example,

  • What is the physical quantity to be measured?
  • Which transducer principle can be used to measure this quantity?
  • What accuracy is required for this measurement?

The first question can be answered by determining the type and range of the measurand. An appropriate answer to the second question requires that the input and output characteristics of the transducer be compatible with the recording or measurement system. In most cases, these two questions can be answered readily, implying that the proper transducer is selected simply by the addition of an accuracy tolerance. In practice, this is rarely possible due to the complexity of the various transducer parameters that affects the accuracy. The accuracy requirements of the total system determine the degree to which individual factors contributing to accuracy must be considered. Some of these factors are:

  1. Fundamental transducer parameters: type and range of measurand, sensitivity, excitation
  2. Physical conditions: mechanical and electrical connections, mounting provisions, corrosion resistance.
  3. Ambient conditions: nonlinearity effects, hysteresis effects, frequency response, resolutions
  4. Environmental conditions: temperature effects, acceleration, shock and vibration
  5. Compatibility of the associated equipment: zero balance provisions, sensitivity tolerance, impedance matching, insulation resistance.

Categories 1 and 2 are basic electrical and mechanical characteristics of the transducer. Transducer accuracy, as an independent component, is contained in categories 3 and 4. Category 5 considers the transducer’s compatibility with its associated system equipment.

The total measurement error in a transducer-activated system may be reduced to fall within the required accuracy range by the following techniques:

  1. Using in-place system calibration with corrections performed in the data reduction
  2. Simultaneously monitoring the environment and correcting the data accordingly.
  3. Artificially controlling the environment to minimize possible errors.

Some individual errors are predictable and can be calibrated out of the system. When the entire system is calibrated, these calibration date may then be used to correct the recorded data. Environmental errors can be corrected by data reduction if the environmental effects are recorded simultaneously with the actual data. Then the data are corrected by using the known environmental characteristics of the transducers. These two techniques can provide a significant increase in system accuracy.

Another method to improve overall system accuracy is to control artificially the environment of the transducer. If the environment of the transducer can be kept unchanged, these errors are reduced to zero. This type of control may require either physically moving the transducer to a more favorable position or providing the required isolation from the environment by heater enclosure, vibration isolation, or similar means.

Thursday, April 21, 2011

Time base generator

It has been mentioned earlier that the time domain oscilloscopes require a sweep generator that is linear with time for the x-axis display. The motion of spot on the screen from extreme left to extreme right is called sweep.

The generator which generates a waveform which is responsible for the movement of spot on screen horizontally is called time base generator or sweep generator. The sweep circuit along with the display gating functions is called time bases.

The linear sweep moves the spot from left to right while the movement of spot from right to left is not visible. This portion of waveform generated by time base is called flyback or retrace. During this time, the cathode ray tube is blanked.

The time base generator also controls the rate at which the spot moves, across the screen. This rate is to be adjusted from front panel control.

Why sweep generator is called time base generator?


All the time dependent waveforms need x-axis to be calibrated as time axis. The sweep generator produces the movement of spot on screen such that it acts as a time axis or time base for the waveforms to be displayed. Hence the sweep generator is also called time base generator.

The two front panel controls which are used to control rate and duration of time base waveform are time/division and time variable control.

Note: the trigger circuit ensures that the horizontal sweep starts at the same point of the vertical input signal.

Requirements of time base:


The time base requirements are:


  1. Sweep time variations from 10nsec to 5 sec per division.
  2. Time accuracy is better than 3%.
  3. Linearity better is better than 1% across the cathode ray tube.
  4. Ten times expansion in the horizontal amplifier which allows 1 nsec per division displays for very high speed transients.
  5. The speed of the spot should be constant across the entire screen.
  6. The spot should be invisible while tracing from right to left and should be visible only from left to right.


Basic principle of time base generator


The basic sweep generator uses the charging characteristics of a capacitor to generate linear risetime voltages. Linearly increasing ramp which becomes zero with very short duration of time ensures that the spot is visible from left to right and invisible from right to left.

The image below shows a capacitor charged from a constant current source.

When switch S1 is closed, S2 is open and capacitor charges to produce linear ramp at the output. The sweep rate can be controlled by changing the value of capacitor or charging current.

Reaching to the maximum value of ramp voltage, the switch S2 is closed and S1 is open. Thus capacitor gets discharged through the resistance R. this is called flyback or retrace. The time t1 is called sweep time. The circuit is a sort of relaxation oscillator which generates saw tooth waveform. But this circuit has less accuracy. The bootstrap techniques allow much greater linearity but he techniques are much more costly.

During the sweep time, the spot moves from left to right. During retrace, the screen is blanked and spot comes back to its starting level but its movement from right to left is invisible.
Practically a trigger circuit is associated with the time base generator. This circuit generates a trigger pulse which activates the time base generator to produce a ramp.

Now when one cycle, sweep and retrace is completed then time base generator takes certain time to start the next cycle. This time is divided into two types as hold off time and waiting time.

Hold off time:


Through the tigger circuit applies the pluse immediately after completion of cycle, the time base generator takes some time to start the ramp. This time is required to stabilize the flyback circuitry. This time is called hold off time.

Waiting time:


Now when trigger pulse is generated b trigger circuit, the pulse has to cross certain reference level so as to activate the time base generator. This reference level is called trigger threshold.

Now after the end of hold off period, though circuit is ready, due to trigger threshold crossing, the pulse takes some time to activate time base generator. So time required by the triggering pulses to cross the trigger threshold is called waiting time.

Tuesday, April 19, 2011

Fluorescent Screen of CRT

The light produced by the screen does not disappear immediately when bombardment by electron ceases, i.e., when the signal becomes zero. The time period for which the trace remains on the screen after the signal becomes zero is known as persistence. The persistence may be as short as a few microseconds, or as long as tens of seconds or even minutes.

Medium persistence traces are mostly used for general purpose applications.

Long persistence traces are used in the study of transients. Long persistence helps in the study of transients since the trace is still seen on the screen after the transient has disappeared.

Short persistence is needed for extremely high speed phenomena.

The screen is coated with fluorescent material called phosphor which emits light when bombarded by electrons. There are various phosphors available which differ in color, persistence and efficiency.

One of the common phosphor is willemite, which is zinc, orthosilicate, ZnO+SiO2, with traces of manganese. This produces the familiar greenish trace. Other useful screen materials include compounds of zinc, cadmium, magnesium and silicon.

The kinetic energy of the electron beam is converted into both light and heat energy when it hits the screen. The heat so produced gives rise in phosphor burn which is damaging and sometimes destructive. This degrades the light output of phosphor and sometimes may cause complete phosphor destruction. Thus the phosphor must have high burn resistance to avoid accidental damage.

Similarly the phosphor screen is provided with an aluminum layer called aluminizing the cathode ray tube. This is shown in image below.


These Aluminizing layers serve three functions:



  1. To avoid buildup of charges on the phosphor which tend to slow down the electrons and limits the brightness.
  2. It serves as a light scatter. When the beam strikes the phosphor with aluminized layer, the light emitted back into the tube is reflected back towards the viewer which increases the brightness.
  3. The aluminum layer acts as a heat sink for the phosphor and thus reduces the chances of the phosphor burning.


Phosphor screen characteristics



  1. Many phosphor materials having different excitation times and colors as well as different phosphorescence times are available.
  2. The type P1, P2, P11 or P31 are the short persistence phosphors and are used for the general purpose oscilloscopes.
  3. Medical oscilloscopes require longer phosphor decay and hence phosphors like P7 and P39 are preferred for such applications
  4. Very slow displays like radar require long persistence phosphor to maintain sufficient flicker free picture. Such phosphors are P19, P26 and P33.
  5. The phosphors P19, P26, P33 have low burn resistance. The phosphors P1, P2, P4, P7, P11 have medium burn resistance while P15,P31 have high burn resistance.


Why P31 is commonly used?


Out of these varieties, the materials P31 is used commonly for general purpose oscilloscopes due to following characteristics:


  1. It gives color to which human eye response is maximum.
  2. It gives short persistence required to avoid multiple image display.
  3. It has high burn resistance to avoid the accidental damage.
  4. Its illumination level is high.
  5. It provides high writing speed.


Note: the light output of a fluorescent screen is proportional to the number of bombarding electrons, i.e., to the beam current.

Saturday, April 16, 2011

Cathode Ray Tube - Deflection system

This Post Deflection system of CRT is a continuation of my previous post Electron Gun of CRT. This post completely covers the Deflection system of the CRT. When the electron beam is accelerated it passes through the deflection system, with which beam can be positioned anywhere on the screen.

The deflection system of the cathode-ray-tube consists of two pairs of parallel plates, referred to as the vertical and horizontal deflection plates. One of the plates in each set is connected to ground (0 V). to the other plate of each set, the external deflection voltage is applied through an internal adjustable gain amplifier stage. To apply the deflection voltage externally, an external terminal, called the y input or the x input, is available.

As shown in the image below, the electron beam passes through these plates. A positive voltage applied to the y input terminal (V y) causes the beam to deflect vertically upward due to the attraction forces, while a negative voltage applied to the y-input terminal will cause the electron beam to deflect vertically downward, due to the repulsion forces.

Similarly, a positive voltage applied to X-input terminal(V x) will cause the electron beam to deflect horizontally towards the right; while a negative voltage applied to the X-input terminal will cause the electron beam to deflect horizontally towards the left of the screen. The amount of vertical or horizontal deflection is directly proportional to the corresponding applied voltage.

When the voltages are applied simultaneously to vertical and horizontal deflecting plates, the electron beam is deflected due to the resultant of these two voltages.

The face of the screen can be considered as an X-Y plane. The (X,Y) position of the beam spot is thus directly influenced by the horizontal and the vertical voltages applied to the deflection plates Vx and Vy respectively.

The horizontal deflection (X) produced will be proportional to the horizontal deflecting voltage, Vx, applied to X-input.

X = KxVx

Where, Kx is constant of proportionality.

The deflection produced is usually measured in cm or as number of division , on the scale, in the horizontal direction.

Then Kx = x/Vx where Kx expressed as cm/volt or division/volt, is called horizontal sensitivity of the oscilloscope.

Similarly, the vertical deflection (y) produced will be proportional to the vertical deflecting voltage, Vy, applied to the y-input.
Y= KyVy

Ky=y/Vy and Ky, the vertical sensitivity, will be expressed as cm/volt, or division/volt.

The schematic arrangement of the vertical and the horizontal plates controlling the position of the spot on the screen is shown in the figure.
deflection-system-crt

The values of vertical and horizontal sensitivities are selectable and adjustable through multi positional switches on the front panel that controls the gain of the corresponding internal amplifier stage. The bright spot of the electron beam can thus trace (or plot) the X-Y relationship between the two voltages, Vx and Vy.

Monday, April 11, 2011

Cathode Ray Tube - Electron Gun

This post is the continuation of the previous post on cathode ray tube.

The electron gun section of the cathode ray tube provides a sharply focused electron beam directed towards the fluorescent-coated screen. This section starts from thermally heated cathode, emitting the electrons. The control grid is given negative potential with respect to cathode. This grid controls the number of electrons in the beam, going to the screen.

The momentum of the electrons (their number * their speed) determines the intensity, or brightness, of the light emitted from the fluorescent screen due to the electron bombardment. The light emitted is usually of the green color. Because the electrons are negatively charged, a repulsive force is created by applying a negative voltage to the control grid (in CRT, voltages applied to various grids are stated with respect to cathode, which is taken as common point). This negative control voltage can be made variable.
electron gun inside crt

Note: a more negative voltage results in less number of electrons in the beam and hence decreased brightness of the beam spot.

Since the electron beam consists of many electrons, beam tends to diverge. This is because the similar (negative) charges on the electron repel each other. To compensate for such repulsion forces, an adjustable electrostatic field is created between two cylindrical anodes called the focusing anodes.

Note: the variable positive voltage on the second anode is used to adjust the focus or sharpness of the bright beam spot.

The high positive potential is also given to the preaccelerating anodes and accelerating anodes, which results into the required acceleration of the electrons.

Both focusing and accelerating anodes are cylindrical in shape having small openings located in the center of each electrode, co-axial with the tube axis. The preaccelerating and accelerating anodes are connected to a common positive high voltage which varies between 2 kV to 10 kV. The focusing anode is connected to a lower positive voltage of about 400V to 500V.

In the next post I will explain about the deflection system of cathode ray tube.

Saturday, April 9, 2011

Cathode Ray Tube (CRT)

The cathode ray Tube (CRT) is the heart of the C.R.O. the CRT generates the electron beam, accelerates the beam, deflects the beam and also has a screen where beam becomes visible as a spot. The main parts of the CRT are:

  1. Electron Gun
  2. Deflection system
  3. Fluorescent screen
  4. Glass tube or envelope
  5. Base

A schematic diagram of CRT, showing its structure and main components is shown in the figure below.

cathode ray tube


Since I want to explain each and every parts of the CRT in detail, I will spilt this topic into 3 parts.

Thursday, March 24, 2011

Temperature compensation and Error in PMMC:

The basic pmmc instrument is sensitive to the temperature. The magnetic field strength and spring tension decrease with increase in temperature. The coil resistance increases with increase in the temperature. Thus pointer reads low for a given current. The meter tends to read low by approximately 0.2% per Celsius rise in the temperature. Hence the temperature compensation is provided by appropriate use of series and shunt resistance of copper and manganin.

Temperature compensation


The simple temperature compensation circuit for PMMC uses a resistance in series with a movable coil, as shown in the figure. The resistor is called swamping resistor. It is made up of manganin having practically zero temperature coefficients, combined with copper in the ratio of 20/1 or 30/1.
temperature compensation for pmmc

The resultant resistance of the coil and the swamping resistor increases slightly as temperature increases, just enough to compensate the change in springs and magnet due to temperature. Thus the effect of temperature is compensated.

More complicated but complete cancellation of temperature effects can be obtained by using the swamping resistors in series and parallel combination as shown in figure.

advanced temperature compensation for pmmc


In this circuit, by correct proportioning of copper and manganin parts, complete cancellation of the temperature effects can be achieved.

Errors in PMMC Instrument


The basic sources of error in PMMC instruments are friction, temperature and aging of various parts. To reduce the frictional errors ratio of torque to weight is made very high.

The most serious errors are produced by the heat generated or by changes in the temperature. This changes the resistance of the working coil, causing large errors. In case of voltmeters, a large series resistance of very low temperature coefficient is used. This reduces the temperature errors.

The aging of permanent magnet and control springs also cause errors. Opposite errors in PMMC is caused by weakening of magnet and spring cause. The weakening of magnet causes less deflection while weakening of control springs cause large deflection, for a particular value of current. The proper use of material and pre-ageing during manufacturing can reduce the errors due to weakening of control springs.

Wednesday, March 23, 2011

Permanent magnet moving coil instruments (PMMC)


The permanent magnet moving coil instruments are most accurate type for direct current measurements. The action of these instruments is based on the motoring principle. When a current carrying coil is placed in the magnetic field produced by permanent magnet, the coil experiences a force and moves. As the coil is moving and the magnet is permanent, the instrument is called permanent magnet moving coil instrument. This basic principle is called D’Arsonval principle. The amount of force experienced by the coil is proportional to the current passing through the coil.

The PMMC instrument is shown in the below images.

permanent magnet moving coil


The moving coil is either rectangular or circular in shape. It has number of turns of fine wire. The coil is suspended so that it is free to turn about its vertical axis. The coil is placed in uniform, horizontal and radial magnetic field of a permanent magnet in the shape of a horse-shoe. The iron core is spherical if coil is circular and is cylindrical if the coil is rectangular. Due to iron core, the deflecting torque increase, increasing the sensitivity of the instrument.

The controlling torque is provided by two phosphor bronze hair springs.

The damping torque is provided by eddy current damping. It is obtained by movement of aluminum former, moving in the magnetic field of the permanent magnet.

The pointer is carried by the spindle and it moves over a graduated scale. The pointer has light weight so that it deflects rapidly. The mirror is placed below the pointer to get the accurate reading by removing the parallax. The weight of the instrument is normally counter balanced by the weights situated diametrically opposite and rapidly connected to it. The scale markings of the basic d.c PMMC instruments are usually linearly spaced as the deflecting torque and hence the pointer deflections are directly proportional to the current passing through the coil.

The top view of PMMC instrument is shown in the below image.
pmmc top view

Advantages of PMMC


The various advantages of PMMC instruments are,

It has uniform scale.
With a powerful magnet, its torque to weight ratio is very high. So operating current of PMMC is small.
The sensitivity is high.
The eddy currents induced in the metallic former over which coil is wound, provide effective damping.
It consumes low power, of the order of 25 W to 200 mW.
It has high accuracy.
Instrument is free from hysteresis error.
Extension of instrument range is possible.
Not affected by external magnetic fields called stray magnetic fields.

Disadvantages of PMMC


The various disadvantages of PMMC instruments are ,

PMMC is Suitable for direct current measurement only.
Ageing of permanent magnet and the control springs introduces the errors.
The cost is high due to delicate construction and accurate machining.
The friction is due to jewel-pivot suspension.